140
Using ProLink
®
Software with Micro Motion
®
Transmitters
Calibrate Menu
continued
Density calibration
2. Perform a two-point, one-point, or three-point density calibration:
• To perform a two-point density calibration for an RFT9739, follow
the procedures in steps 3 and 4.
• If performing a one-point density calibration, follow either step 3
or step 4.
• The third-point density calibration procedure is described in
step 5.
3. To perform the low-density calibration:
a. Fill the sensor with a low-density fluid, such as air.
b. If possible, shut off the flow. Otherwise, pump the fluid through
the sensor at the lowest flow rate allowed by the process.
c. Use any established method to derive an accurate density, in
grams per cubic centimeter, for the fluid at line conditions. If air is
the low-density calibration fluid, a value from
Table 14-2
can be
used for the density.
d. Open the Calibrate menu, then choose Density. At the cascading
menu, choose First Point (Air) to open the Density Point 1 dialog
box, shown in
Figure 14-4
.
e. Enter the line-condition density,
in grams per cubic centimeter
,
into the Enter Density text box, then choose Do Cal. The
transmitter measures the tube period, corrects it to 0°C, and
stores the value. The indicator in the dialog box blinks ON and
OFF while density calibration is in progress. When the indicator
stops blinking, the low-density calibration is complete.
Figure 14-4. Density Point 1 Calibration dialog box for RFT9739