Using ProLink
®
Software with Micro Motion
®
Transmitters
Configure Menu: Transmitter Information
continued
Burst control for RFT9739 and RFT9712/RFT9729
Modbus pressure compensation
The Modbus-compatible host controller downloads pressure values to
the RFT9739 by writing an integer that represents gauge pressure to
input register 40007, or by writing an IEEE 754 floating point value that
represents pressure to register pair 20257-20258.
The user must configure the RFT9739 for pressure compensation, as
described in
Section 8.7
, page 86.
Control output
The RFT9739 control output can be configured to indicate pressure
input failure. In this configuration, if an input failure is indicated, one of
the following has occurred:
The mA or digital input from the pressure transmitter is out of range
(less than 2 mA or greater than 22 mA).
The pressure device failed to respond to a HART command.
11.4 Burst control for RFT9739
and RFT9712/RFT9729
Use the Burst option buttons to determine whether the RFT9739,
RFT9712, or RFT9729 will communicate in burst mode under HART
protocol. The IFT9701 and IFT9703 do not support burst mode data
transmission.
Before switching burst mode ON, configure the transmitter to use HART
protocol (see the transmitter instruction manual), then use a Model 275
HART Communicator or AMS with the transmitter to configure the burst
mode.
If ON is chosen, the transmitter bursts data at regular intervals via the
fixed 4 mA or RS-485 wiring loop. The interval varies according to the
baud rate. At 1200 baud, the host receives two updates every 3.7
seconds.
The RFT9739 can burst HART commands 1, 2, 3, 33, and 110. The
default command after a master reset is 2.
The RFT9712 and RFT9729 can burst HART commands 1, 2, 3, and
110. The default command after a master reset is 1.
HART commands are listed in
Table 11-2
.
On a multidrop network, only one transmitter at a time may be in
burst mode. Burst mode can be enabled regardless of the transmitter
address.
When using the Bell 202 standard, the transmitter primary variable
milliamp output will become fixed if the transmitter has an address
other than zero (0).
For RFT9712 and RFT9729 transmitters, when the display is opened,
neither button will be selected, as shown in
Figure 11-3
, because the
ProLink software cannot predetermine the RFT9712 or RFT9729 burst
state.