Using ProLink
®
Software with Micro Motion
®
Transmitters
Getting Started
continued
Connecting to the transmitter
Transmit light on PC Interface adaptor
does not flash when trying to connect
to the transmitter
Incorrect computer communication port Change comm ports with the ProLink
®
setup program
Check cable connections between the
computer and the PC Interface adaptor
Computer communication port (COMM1
or COMM2) is being used by another
program or device, such as a mouse,
fax, or modem
Change comm ports with the ProLink
setup program
Disable device and remove other
conflicting comm drivers
IRQ is being shared by ProLink and
another program or device, such as a
mouse, fax, or modem, which is using
communication port 3 (COMM3) or 4
(COMM4)
Disable device and remove other
conflicting comm drivers
Windows 3.1 and ProLink time source in
conflict
Change the time source from Real Time
clock to Interval Timer with the ProLink
setup program
Transmit light on PC Interface adaptor
remains OFF, and the receive light
flashes while trying to connect to
transmitter
Using wrong type of cable between the
computer and the PC Interface adaptor
Use a straight through RS-232 cable
from the computer to the PC Interface
adaptor
Transmit light on PC Interface adaptor
flashes while trying to connect to
transmitter using the Bell 202 physical
layer
Conflict between physical layer settings Ensure that transmitter and position of
selector switch on PC Interface adaptor,
are all set to Bell 202
Improper wiring between PC Interface
adaptor and transmitter
Ensure proper Bell 202 wiring (see
“Wiring to the transmitter” on page 7)
Incorrect load resistance in wiring loop Ensure proper load in Bell 202 wiring
loop (see “Wiring to the transmitter” on
page 7)
Incorrect communication parameters Bell 202 requires HART
®
protocol at
1200 baud, 1 stop bit, and odd parity
(see “Communication options” on page
22, to change settings)
Noise on mA loop from an external
source
Ensure proper resistance in the Bell 202
wiring loop (see “Wiring to the
transmitter” on page 7)
RFT9739 not properly configured for
Bell 202
Change RFT9739 settings (see the
instruction manual that was shipped with
the transmitter)
RFT9712 or RFT9729 not properly
configured for Bell 202
See the RFT9712 or RFT9729
instruction manual to verify:
The communications jumper located
on the processor board in the
RFT9712 or RFT9729 is set to 268
The RFT9712 or RFT9729 has
transmitter software version 5.0 or
higher
Incompatible communication settings
between IFT9701/IFT9703 and ProLink
program
Ensure ProLink settings are configured
for HART protocol, 1200 baud, 1 stop bit,
250-600 ohm resistance and odd parity
(see “Communication options” on page
22)
Table 2-2. Troubleshooting the "cannot find" message
Status Cause Corrective Action