Using ProLink
®
Software with Micro Motion
®
Transmitters
57
View Menu: Status
continued
Critical indicators
Configure Menu:
Transmitter Outputs
Configure Menu:
Transmitter Information
Configure Menu: EventsView Menu: Status
Configure Menu:
Characterize
Configure Menu:
Transmitter Variables
"Not Configured"
If a master reset is performed on the transmitter, Not Configured
switches ON to indicate the flowmeter requires complete
characterization and reconfiguration. A master reset returns all
transmitter options to their default factory values.
• To perform a master reset, refer to the transmitter instruction manual.
• To characterize the flowmeter, open the Configure menu, then
choose Characterize. For more information about characterization,
see
Chapter 8
(page 71).
Transmitter failure
indicators
If one or more of the following indicators switches ON, a transmitter
failure has occurred:
• "Transmitter Failure"
• "(E)EPROM Checksum"
• "RAM Error"
•"RTI Failure"
• "0 Deg Cal Failure"
• "200 Deg Cal Failure"
If any of these errors occur, contact the Micro Motion Customer
Service department.
Sensor failure and
overrange indicators
If a sensor failure occurs, if the sensor cable is faulty, or if measured flow,
measured temperature, or measured density goes outside the sensor
limits, one or more of the following indicators switches ON:
• "Sensor Failure"
•"RTD Failure"
• "Drive Overrange"
• "Input Overrange"
• "Temperature Overrange"
• "Mass Flow Overrange"
• "Volume Flow Overrange"
• "Density Overrange"
CAUTION
During troubleshooting, the transmitter could
produce inaccurate flow signals.
• Set control devices for manual operation before
troubleshooting the flowmetering system.
• If terminal blocks or wiring are disconnected from the
transmitter during troubleshooting, cycle power to the
transmitter after reconnecting terminal blocks or
wiring.