134
Using ProLink
®
Software with Micro Motion
®
Transmitters
Calibrate Menu
continued
Auto zero
4. Open the Calibrate menu, then choose Auto Zero. The Flow
Calibration dialog box will be displayed, as depicted in:
•
Figure 14-1
for an RFT9739
•
Figure 14-2
for an IFT9701 or IFT9703
•
Figure 14-3
for an RFT9712 or RFT9729
5.
Make sure flow through the sensor is completely stopped,
then
choose Zero. The transmitter begins zeroing.
• The indicator in the lower right corner of the Flow Calibration
dialog box blinks to indicate transmitter zeroing is in progress.
• Transmitters with a display: the message screen reads CAL IN
PROGRESS or ZERO during transmitter zeroing.
• Field-mount transmitters: the diagnostic LED remains ON during
transmitter zeroing.
To end the zero operation before its completion, choose Stop.
6. When transmitter zeroing is completed, the Zero In Progress
indicator stops blinking. Choose Done to exit the Flow Calibration
dialog box.
• Transmitters with a display: the message screen reads ERROR
CLEARED.
• Field-mount transmitters: the diagnostic LED blinks ON once per
second to indicate normal operation.