Using ProLink
®
Software with Micro Motion
®
Transmitters
Configure Menu: Characterize
continued
Flow calibration factor
Sensor or transmitter
replaced in the field
If the sensor or transmitter is replaced in the field, the user must
re-characterize the transmitter by programming the new sensor flow
calibration factor into the transmitter. (For an RFT9739, IFT9701,
IFT9703, RFT9712, or RFT9729 that is shipped without a sensor, the
default value — which
must
be changed — is 1.00005.13).
To characterize the transmitter for flow measurement:
Enter the flow calibration factor (eight digits and two decimal points)
from the sensor serial number tag — this is the preferred method;
or
Enter the flow calibration factor (eight digits and two decimal points)
from the calibration certificate that was shipped with the sensor.
Field flow-calibration
Flow calibration is performed at the factory to NIST (National Institute of
Standards and Technology) standards. If, after characterizing the
transmitter for flow as described above, the meter fails to perform within
the accuracy specifications provided by Micro Motion, it might be
necessary to perform a field flow-calibration or, for a Version 3.0, 3.5, or
3.6 RFT9739, to adjust the meter factor (see
Chapter 13
).
If a field flow-calibration is performed, and the transmitter is an RFT9739
using a pressure input for pressure compensation, the user should enter
the calibration pressure in the FlowCal Pressure text box. Enter the
value in psi. For more information on pressure compensation, see
Section 8.7
, page 86.
In the flow calibration procedure, a batch of fluid is run through the
sensor, then the weighed amount of fluid in the batch is compared with
the measured amount of fluid indicated in the Totalizer Control dialog
box. To perform a flow calibration, follow these steps:
1. Set process control devices for manual operation.
2. For an IFT9701, IFT9703, RFT9712, or RFT9729, if the application
requires volume flow measurement, choose a mass flow unit for the
flow calibration, then choose a volume flow unit for the application
after the calibration is complete. The flow calibration factor will be the
same regardless of the mass flow unit that is chosen.
To select a mass flow unit:
a. Open the Configure menu, then choose Transmitter Variables.
b. When the Configure Transmitter Variables dialog box appears,
open the Units list box under Mass/Volume Flow (for an IFT9701
or IFT9703) or Mass Flow (for an RFT9712 or RFT9729) to select
a standard engineering mass flow unit.
For more information about configuring flow units, see
Section 9.2
,
page 93.