66
Using ProLink
®
Software with Micro Motion
®
Transmitters
View Menu: Status
continued
Informational indicators
"Raw Flow Overflow" and
"Raw Elec. Zero Overflow"
Raw Flow Overflow and Raw Elec Zero Overflow switch ON to indicate
the transmitter cannot calculate flow. If there are no open circuits in the
sensor wiring, the sensor is being subjected to mounting stress or
external noise.
If Raw Flow Overflow or Raw Elec Zero Overflow switches ON, eliminate
mounting stress on the sensor, and isolate the sensor from vibration and
mechanical noise.
7.5 Informational indicators
Informational indicators switch on to indicate the following changes in
operating states:
• Transmitter start-up, calibration, zeroing, and testing
• Polling address other than 0 assigned to the transmitter for HART
communication in a Bell 202 multidrop network
• Cancel chosen during an output calibration or test
• Transmitter configured to send data in burst mode under HART
protocol
• An event assigned to an RFT9739 output achieves the setpoint
• An interruption or error occurs in analog input
Table 7-8
, page 68 summarizes conditions indicated by informational
indicators.
Example:
Given:
An RFT9739 is configured to indicate flow rate in pounds per
minute (lb/min). The frequency/pulse output has been scaled so that
100 Hz = 1 lb/min. The frequency/pulse output goes out of range at
15 kHz.
Problem:
When the process flow rate exceeds 150 lb/min, the
frequency/pulse output is saturated.
Solution:
Increase the flow rate represented by the output at 15 kHz by
rescaling the frequency/pulse output so that 100 Hz = 2 lb/min (the
output will be saturated at 300 lb/min).