Using ProLink
®
Software with Micro Motion
®
Transmitters
Calibrate Menu
continued
Programming auto zero for RFT9739
Diagnosing zeroing failure
If zeroing fails, indicators in the Status window switch ON. To view the
Status window, open the View menu, then choose Transmitter Status.
The Calibration Failure indicator switches ON
If the transmitter is an RFT9739, RFT9712, or RFT9729, the Zero
Too Noisy, Zero Too High, or Zero Too Low might also switch ON.
The most common sources of zeroing failure are:
Flow of fluid through sensor during zeroing
Partially empty sensor flow tubes
An improperly mounted sensor
To clear a zeroing error, re-zero the transmitter after correcting the
problem, or cycle power to the transmitter to abort the procedure and
return to the previously established zero. For an RFT9739, adjust the
programmable auto zero parameters before re-zeroing. See
Section 14.3
.
14.3 Programming auto zero for
RFT9739
If Calibration Failure switches ON in the Status window, auto zero has
failed. If the Zero Too Noisy, Zero Too Low, or Zero Too High indicator is
also ON, and the zeroing procedure was performed on an RFT9739, the
auto zero parameters can be reprogrammed.
During the zeroing process, the RFT9739 measures the time shift (the
time between signals from the left and right sensor pickoff coils) for each
measurement cycle, computes the average time shift per cycle, then
derives the standard deviation of the average time shift over the zero
time.
If zeroing is too noisy, the user can program a standard deviation limit
to account for mechanical noise from vibrating pumps or other
equipment. Mechanical noise can cause zero failure by interfering
with signals from the sensor.
Zero time is the number of measurement cycles required for
transmitter zeroing, where one measurement cycle equals two tube
periods. The default zero time is 2048 cycles (approximately 40
seconds at a tube frequency of 100 Hz). A longer zero time might
improve the accuracy of the zeroing procedure by increasing the
number of measurement cycles.
Convergence limit
To reprogram the convergence (standard deviation) limit, enter the
desired value into the Convergence Limit text box (see
Figure 14-1
,
page 135). Record the convergence limit in the appropriate Flowmeter
Calibration Record in
Appendix F
, page 189.