Micro Motion RFT9709 Instrukcja Obsługi Strona 152

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 214
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 151
138
Using ProLink
®
Software with Micro Motion
®
Transmitters
Calibrate Menu
continued
Density calibration
14.4 Density calibration
Density calibration adjusts the factors that are used by the transmitter to
calculate density. Fluid density, calculated by the transmitter, is inversely
proportional to the square of the sensor tube frequency. The ProLink
program supports three types of density calibration: one-point, two-point
(or dual-point), and third-point.
During two-point density calibration, the user commands the transmitter
to measure the tube period when the flow tubes contain a fluid with a
reference low density (usually air) and when the flow tubes contain a
fluid with a reference high density (usually water). The transmitter then
calculates the density of the process fluid from the following equation:
Where:
D = Calculated density at measured temperature t
D
1
= Reference density of low-density calibration fluid at calibration
temperature t
1
D
2
= Reference density of high-density calibration fluid at
calibration temperature t
2
τ
2
= Measured tube period squared at measured temperature t,
corrected to 0°C
τ
1
2
= Tube period squared at D1, corrected to 0°C
τ
2
2
= Tube period squared at D2, corrected to 0°C
Two-point density calibration is preferably performed under zero flow
conditions.
One-point density calibration is an option when two-point calibration is
not possible. However, because one-point calibration is performed with
only one reference density, two-point calibration is recommended for
increased accuracy.
Third-point density calibration is available only for the RFT9739. As the
flow rate approaches the maximum flow rate of the sensor, the angular
momentum of the fluid can alter the tube period, which causes an
increase in the density value measured by the sensor. An increase in the
flow rate causes a proportional increase in the angular momentum of the
fluid. A third-point density calibration accounts for the effect of flow on
the tube period at high flow rates.
Third-point density calibration is desirable if the process exceeds or
often approaches the flow rate listed in
Table 14-1
.If the process
remains below this rate, calibrating a third density point is unnecessary,
since angular momentum of the fluid will have a minimal effect on the
flow tube frequency. Use the data listed in
Table 14-1
to determine
whether third-point density calibration is desirable.
DD
2
D
1
()
τ
2
τ
1
2
τ
2
2
τ
1
2
----------------
×
D
1
+=
Przeglądanie stron 151
1 2 ... 147 148 149 150 151 152 153 154 155 156 157 ... 213 214

Komentarze do niniejszej Instrukcji

Brak uwag