Using ProLink
®
Software with Micro Motion
®
Transmitters
139
Calibrate Menu
continued
Density calibration
Applications Menu Help Menu
Configure Menu:
Meter Factors
Calibrate Menu Test Menu
Table 14-1. Minimum flow rate for third-point calibration
Density calibration for
RFT9739
During density calibration, write values in the appropriate Flowmeter
Calibration Record, shown in
Appendix F
, page 189. Follow these steps
to perform a density calibration for an RFT9739:
1. Density calibration requires reading and entering density values in
grams per cubic centimeter. To change the density unit of measure:
a. Open the Configure menu, then choose Transmitter Variables.
b. In the Configure Transmitter Variables dialog box, open the
density units list box, then select g/cc as the density unit.
c. Click OK.
Sensor model
Minimum flow rate
lb/min kg/h
ELITE
®
CMF010 2.5 69
CMF025 27 720
CMF050 86 2350
CMF100 280 7575
CMF200 1270 34,540
CMF300 4390 119,600
CMF400 15,000 409,000
T- S e ri es
T075 500 13,630
T100 1100 29,990
T150 3500 95,430
BASIS
®
F200 2315 63,045
All other F sensors K3 density calibration not necessary
Model D
D6 0.8 25
D12 4.5 125
D25 18 485
D40 stainless steel 33 900
D40 Hastelloy
®
C-22 52 1395
D65 115 3060
D100 405 11,010
D150 1140 31,050
D300 2705 73,660
D600 9005 245,520
Model DH
All DH sensors K3 density calibration not necessary
Model DL
DL65 115 3075
DL100 325 8,780
DL200 1210 32,950
Model DT
DT65 150 4040
DT100 315 8460
DT150 580 15,780